| GENERAL SPECIFICATIONS |
 |
| Performance |
 |
 |
 |
| Sensor Classification |
|
Colormetric with glass flowcell and LED source |
 |
 |
 |
| Application |
|
Demineralized,boiler,potable,surface and waste water |
 |
 |
 |
| Power Requirements |
|
12V |
 |
 |
 |
| Power Supply |
|
110-120V or 220-240V 50/60 Hz, 100 VA |
 |
 |
 |
| Humidity |
|
Up to 90% not condensable |
 |
 |
 |
| Ambient Temperature Range |
|
10-45º C analyzer (50-113º F) |
 |
 |
 |
| Range |
|
Silica: 0-500 ppb, 0-5000 ppb, 0-200 ppm*
Phosphate: 0-2.4 ppm,0-7 ppm, 0-20 ppm, 0-200 ppm*
Hydrazine, Copper:0-100 ppb, 0-1 ppm*
Ammonia: 0-750 ppb, 0-7.5 ppm*
Ethylene Glycol: 0-15 ppm, 0-150 ppm*
*By Auto Dilution |
 |
 |
 |
| Accuracy |
|
Silica: +/– 1 ppb or +/– 2% of reading, whichever is greater
Phosphate, Hydrazine, Ammonia, Copper, Ethylene Glycol: +/- 2% of full scale |
 |
 |
 |
| Unit Dimensions |
|
Height = 15.5" (39.4 cm), Width = 6.25” (15.88 cm), Depth = 22” (55.9 cm) |
 |
 |
 |
| Atmospheric Pressure Range |
|
No Limits |
 |
 |
 |
| Effect of Electromagnetic Fields |
|
EMC tested according to CE compliance |
 |
 |
 |
| Tolerance to Electrostatic Discharges |
|
EMC tested according to CE compliance |
 |
 |
 |
| Response Time (approximate) |
|
Silica: 10 min, Phosphate: 6 min, Hydrazine: 12 min, Ammonia: 12 min, Copper: 9 min, Ethylene Glycol: 10 min |
 |
 |
 |
| Alarms |
|
Concentration,cal error,out of service/sample |
 |
 |
 |
| Reproducibility |
|
Silica: +/– 1 ppb or +/– 2% of reading, whichever is greater
Phosphate, Hydrazine, Ammonia, Copper, Ethylene Glycol: +/– 2% of reading |
 |
 |
 |
| Degree of Protection |
|
IP65 (NEMA 4x |
 |
 |
 |
| Required Maintenance |
|
Replenish reagents,once a year replace tubing |
 |
 |
 |
| Reagent Consumption (approximate) |
|
Silica, Phosphate, Ammonia, Copper: 1 liter each per month
Hydrazine, Ethylene Glycol: 4 liters each per month |
 |
 |
 |
| Reactor Volume |
|
10ml |
 |
 |
 |
| Materials in Contact with Sample |
|
Glass, Silicone, Plexiglass, Stainless Steel AISI 316 |
 |
 |
 |
| CE Certification |
|
Meets low voltage and low electromagnetic compatibility directives |
 |
 |
 |
| Sample Conditioning Requirements |
|
Filtering between 10 and 60 micron, depending on the matrix, needed only to avoid clogging. Internal disposable filters supplied (60 micron). |
 |
 |
 |
| |
|
|
| Sample Delivery Operating Conditions |
 |
 |
 |
| Temperature Range |
|
5-55°C (41-131° F) |
 |
 |
 |
| Turbidity |
|
Not applicable; sample blank correction |
 |
 |
 |
| Color |
|
Not applicable; sample blank correction |
 |
 |
 |
| PH |
|
3-12 |
 |
 |
 |
| |
|
|
| Signal Outputs |
 |
 |
 |
| 4-20 mA, 0-5V |
|
4-20mA (galvanic isolator module available as option) or 0-5V |
 |
 |
 |
| Printer Options |
|
Optional, serial output RS 232 or RS 485 |
 |
 |
 |
| Serial I/O for Signals |
|
Serial data output RS 232 or RS 485 available as option |
 |
 |
 |
| |
|
|
| Operational Calibration |
 |
 |
 |
| Frequency/Intervals |
|
Recommended.: 7 days |
 |
 |
 |
| Single/Multi Point |
|
Multi-Point:zero and mid range |
 |
 |
 |
| Matrix Corrections |
|
Yes, sample blank correction |
 |
 |
 |
| Manual/Automatic |
|
Both |
 |
 |
 |
| |
|
|