µAI 7040 Portable Series - Silica and Phosphate Analyzer

µAI 7040 Portable Series

Silica, Phosphate, Hydrazine, Ammonia, Copper and Ethylene Glycol Analyzers

REVOLUTIONARY PATENTED TECHNOLOGY

Our Colorimetric Analyzers (Aqualyzer®Series 7040) are based on the new patented Loop Flow Analysis (LFA) cycle.

Common Silica and Phosphate Analyzer application points in a boiler system for silica measurement include: make-up water, condensate, boiler feedwater, saturated steam/mainsteam and cooling water. Application points for phosphate include:boiler water and cooling water.

LFA technology involves isolating,manipulating and testing a known volume of sample in a closed loop system. A powerful programmable microprocessor allows for customized program cycles,ensuring accuracy and reliability with every application..

Download product brochure (PDF)

 


µAI 7040 Portable Series - Silica and Phosphate Analyzer

 
FEATURES   BENEFITS
Highest accuracy (e.g.for silica:+/– 1ppb or +/– 2% of reading)   Low reagent and standard consumption
Widest range analysis (e.g.for silica:ranges available from 0.1ppb to 200ppm)   Minimal maintenance required
Single,dual stream or multi-parameters SiO2/PO4   Automatic operation
Short cycle times    
Automatic dilution when above normal concentration range    
Automatic calibration    
Adjustable cycle and calibration frequency    
Ultra-low reagent consumption and smallest size instruments    
No reagent pumps or compressed air requirements    
Grab sample capability    
Last 400 results stored internally    
     
 
GENERAL SPECIFICATIONS
Performance
Sensor Classification   Colormetric with glass flowcell and LED source
Application   Demineralized,boiler,potable,surface and waste water
Power Requirements   12V
Power Supply   110-120V or 220-240V 50/60 Hz, 100 VA
Humidity   Up to 90% not condensable
Ambient Temperature Range   10-45º C analyzer (50-113º F)
Range  

Silica: 0-500 ppb, 0-5000 ppb, 0-200 ppm*
Phosphate: 0-2.4 ppm,0-7 ppm, 0-20 ppm, 0-200 ppm*
Hydrazine, Copper:0-100 ppb, 0-1 ppm*
Ammonia: 0-750 ppb, 0-7.5 ppm*
Ethylene Glycol: 0-15 ppm, 0-150 ppm*
*By Auto Dilution

Accuracy   Silica: +/– 1 ppb or +/– 2% of reading, whichever is greater
Phosphate, Hydrazine, Ammonia, Copper, Ethylene Glycol: +/- 2% of full scale
Unit Dimensions   Height = 15.5" (39.4 cm), Width = 6.25” (15.88 cm), Depth = 22” (55.9 cm)
Atmospheric Pressure Range   No Limits
Effect of Electromagnetic Fields   EMC tested according to CE compliance
Tolerance to Electrostatic Discharges   EMC tested according to CE compliance
Response Time (approximate)   Silica: 10 min, Phosphate: 6 min, Hydrazine: 12 min, Ammonia: 12 min, Copper: 9 min, Ethylene Glycol: 10 min
Alarms   Concentration,cal error,out of service/sample
Reproducibility   Silica: +/– 1 ppb or +/– 2% of reading, whichever is greater
Phosphate, Hydrazine, Ammonia, Copper, Ethylene Glycol: +/– 2% of reading
Degree of Protection   IP65 (NEMA 4x
Required Maintenance   Replenish reagents,once a year replace tubing
Reagent Consumption (approximate)   Silica, Phosphate, Ammonia, Copper: 1 liter each per month
Hydrazine, Ethylene Glycol: 4 liters each per month
Reactor Volume   10ml
Materials in Contact with Sample   Glass, Silicone, Plexiglass, Stainless Steel AISI 316
CE Certification   Meets low voltage and low electromagnetic compatibility directives
Sample Conditioning Requirements   Filtering between 10 and 60 micron, depending on the matrix, needed only to avoid clogging. Internal disposable filters supplied (60 micron).
     
Sample Delivery Operating Conditions
Temperature Range   5-55°C (41-131° F)
Turbidity   Not applicable; sample blank correction
Color   Not applicable; sample blank correction
PH   3-12
     
Signal Outputs
4-20 mA, 0-5V   4-20mA (galvanic isolator module available as option) or 0-5V
Printer Options   Optional, serial output RS 232 or RS 485
Serial I/O for Signals   Serial data output RS 232 or RS 485 available as option
     
Operational Calibration
Frequency/Intervals   Recommended.: 7 days
Single/Multi Point   Multi-Point:zero and mid range
Matrix Corrections   Yes, sample blank correction
Manual/Automatic   Both
     
 
       
 