µAI 4620 and µAI 4625

Conductivity Analyzer

THIS PRODUCT IS DISCONTINUED.

The Conductivity Analyzers µAI-4620 wall mount and µAI-4625 panel mount are user-programmable analyzers designed for use in Utilities, Semi-conductor, Chemical Processing, and Pharmaceutical industries.

 

 
FEATURES   BENEFITS
Ultra-Pure Compensation   Ultra-pure water temperature compensation insures accuracy at conductivities of less than 1 NS/cm.
Wide Measurement Range   Wide range of measurement covers applications from ultra-pure water to liquids with conductivity up to 10,OOONS/cm.
Universal Scaling   Incorporates universal scaling to provide the best resolution for your application requirement. Measurement range can be scaled to NS/cm, pS/m, mS/cm, mS/m, Mohm-cm, TDS and ppm.
m On-Line Diagnostics   Comprehensive diagnostics continuously verify analyzer operation of the entire sensing loop with built-in software protection.

SENSING SYSTEMS
 
 
 
 
FLOW CELL MODEL 2045
SCREW-IN MODEL 2077
SCREW-IN MODEL 2078
WITHDRAWABLE CELL - MODEL 2085
Cell constant
available:
0.1 or 1.0
0.1 or 1.0
0.05 or 0.1
0.05 or 0.1
Type:
Flow-line
Screw type
Screw-in
Withdrawable
Cell Body:
Loaded epoxy resin
Loaded epoxy resin
316 St. Steel
316 St. Steel
Electrode Matl.:
Carbon
Carbon
316 St. Steel
Naval brass & 316 St Steel
Connections:
Threaded NPT
Threaded NPT
Threaded 3/4" NPT
Use with Model 2089 valve assembly
Maximum press.:
100 psi
100 psi
150 psi
150 psi
Maximum temp.:
194°F (90°C)
194°F (90°C)
230°F (110°C)
230°F (110°C)
NOTE: To select the most suitable cell constant for the desired range use the following criteria:
Minimum range 10 x the cell constant
Maximum range 10,000 x the cell constant
e.g. cell constant k=0.1      Minimum range = 10 x 0.1 = LOONS/cm
                                          Maximum range = 10,000 x 0.1 = 1000NS/cm

For ultra-pure water applications (less than LOONS/cm) a cell constant of 0.05 must be used.


SPECIFICATIONS
      

Display
Measured Value:

5 digit, 7 segment back-lit LCD.

Information:
16 character, single line, dot matrix, back-lit LCD.

Ranges:
Programmable 0 to 0.5NS/cm up to 10,OOOUS/cm (with various cell constants).

Scaling:
µS/cm, µS/m, mS/cm, mS/m, Mohm-cm, TDS and ppm.

Accuracy:
±1 f.s.d., ± 1 digit.

Linearity:
±0.2% f.s.d.

Temperature measuring range:
14° to 230°F (-10° to 110°C).

Temperature compensation:
14° to 230°F (-10° to 110°C) automatic.

Temperature coefficient:
Programmable 0 to 1.5%/°F (0 to 3.0%/°C).

Temperature sensor:
Pt100 resistance thermometer.

Reference Temperature:

68°F (20°C) or 77°F (25°C) programmable.

Environmental Data
Operating temperature limits:

-4° to 131 `F (-20` to 55°C).

Storage temperature limits:
-13° to 131 °F (-25° to 55°C).

Operating humidity limits:
Up to 95% RH non-condensing.

Power Supply
Voltage requirements:

100 to 130V, 200 to 260V, 50/60Hz.

Power Consumption:
<6VA a.c.

  Error due to power supply variation:
Less than 2% for +6% - 20% variation from nominal supply voltage.

Insulation Line to ground:
2kV r.m.s.

Outputs and Set Points
Number of relays/set points:

Two.

Relay contacts:
Single pole changeover.
Rating    250V a.c.   250V d.c. max.
              3A a.c.       3A d.c. max.

Loading
       (non-inductive) 750VA   30W max.
       (inductive)        750VA   3W max.

Insulation:
2kV r.m.s. contacts to earth (ground).

Number of set points:
Two.

Set point adjustment:
Programmable.

Set point hysteresis:
± 1 % fixed.

Local set point annunciation:
Red L.E.D.

Retransmission:
Number of retransmission signals:
One fully isolated.

Output current:
0 to 10mA, 0 to 20mA or 4 to 20mA programmable.

Accuracy:
±25% f.s.d. or ±0.5% reading.

Resolution:
0.1 % at 10mA, 0.05 at 20mA.

Maximum load resistance:
750 ohm (20mA max.).

Serial communication:
RS485 (optional extra).
   
Mechanical Data
  Model 4620   Model 4625
Mounting Wall   Panel
Protection NEMA 4X/IP66   NEMA 4X /IP66 front
Dimensions 6.36" W x 8.43" H x 2.68" D   3.78" W x 3.78" H x 7.52" D
Panel cutout     (3.62 (+0.03/-0) x 3.62 (+0.03/-0))
Weight 4 1/2 Ibs.   3 1/4 Ibs.

 
       

 