Spare
Parts
Reagents
Manuals
MSDS
Sheets
Shopping Cart
To
order this complete analyzer, please call our sales office at
(800) 242-7353 |
|
|
µAI
4620 and µAI 4625
Conductivity Analyzer
THIS PRODUCT IS DISCONTINUED.
The Conductivity Analyzers µAI-4620 wall mount and µAI-4625 panel mount are user-programmable analyzers
designed for use in Utilities, Semi-conductor, Chemical Processing, and
Pharmaceutical industries. |
|
|
| FEATURES |
|
BENEFITS |
 |
| Ultra-Pure
Compensation |
|
Ultra-pure
water temperature compensation insures accuracy at conductivities of less
than 1 NS/cm. |
 |
 |
 |
| Wide
Measurement Range |
|
Wide
range of measurement covers applications from ultra-pure water to liquids
with conductivity up to 10,OOONS/cm. |
 |
 |
 |
| Universal
Scaling |
|
Incorporates
universal scaling to provide the best resolution for your application requirement.
Measurement range can be scaled to NS/cm, pS/m, mS/cm, mS/m, Mohm-cm, TDS
and ppm. |
 |
 |
 |
| m On-Line
Diagnostics |
|
Comprehensive
diagnostics continuously verify analyzer operation of the entire sensing
loop with built-in software protection. |
 |
 |
 |
| SENSING
SYSTEMS |
| |
|
|
|
|
|
|
|
|
FLOW
CELL MODEL 2045 |
|
SCREW-IN
MODEL 2077 |
|
SCREW-IN
MODEL 2078 |
|
WITHDRAWABLE
CELL - MODEL 2085 |
 |
 |
 |
Cell
constant
available:
|
0.1
or 1.0 |
|
0.1
or 1.0 |
|
0.05
or 0.1 |
|
0.05
or 0.1 |
 |
 |
 |
| Type: |
Flow-line |
|
Screw
type |
|
Screw-in |
|
Withdrawable |
 |
 |
 |
| Cell
Body: |
Loaded
epoxy resin |
|
Loaded
epoxy resin |
|
316
St. Steel |
|
316
St. Steel |
 |
 |
 |
| Electrode
Matl.: |
Carbon |
|
Carbon |
|
316
St. Steel |
|
Naval
brass & 316 St Steel |
 |
 |
 |
| Connections: |
Threaded
NPT |
|
Threaded
NPT |
|
Threaded
3/4" NPT |
|
Use
with Model 2089 valve assembly |
 |
 |
 |
| Maximum
press.: |
100
psi |
|
100
psi |
|
150
psi |
|
150
psi |
 |
 |
 |
| Maximum
temp.: |
194°F
(90°C) |
|
194°F
(90°C) |
|
230°F
(110°C) |
|
230°F
(110°C) |
 |
 |
 |
NOTE:
To select the most suitable cell constant for the desired range use the
following criteria:
Minimum range 10 x the cell constant
Maximum range 10,000 x the cell constant
e.g. cell constant k=0.1 Minimum range = 10
x 0.1 = LOONS/cm
Maximum
range = 10,000 x 0.1 = 1000NS/cm
For ultra-pure water
applications (less than LOONS/cm) a cell constant of 0.05 must be used. |
| SPECIFICATIONS |
| |
|
|
Display
Measured Value:
5 digit, 7 segment back-lit LCD.
Information:
16 character, single line, dot matrix, back-lit LCD.
Ranges:
Programmable 0 to 0.5NS/cm up to 10,OOOUS/cm (with various cell constants).
Scaling:
µS/cm, µS/m, mS/cm, mS/m, Mohm-cm, TDS and ppm.
Accuracy:
±1 f.s.d., ± 1 digit.
Linearity:
±0.2% f.s.d.
Temperature measuring range:
14° to 230°F (-10° to 110°C).
Temperature compensation:
14° to 230°F (-10° to 110°C) automatic.
Temperature coefficient:
Programmable 0 to 1.5%/°F (0 to 3.0%/°C).
Temperature sensor:
Pt100 resistance thermometer.
Reference Temperature:
68°F (20°C) or 77°F (25°C) programmable.
Environmental
Data
Operating temperature limits:
-4° to 131 `F (-20` to 55°C).
Storage temperature limits:
-13° to 131 °F (-25° to 55°C).
Operating humidity limits:
Up to 95% RH non-condensing.
Power Supply
Voltage requirements:
100 to 130V, 200 to 260V, 50/60Hz.
Power Consumption:
<6VA a.c.
|
|
Error
due to power supply variation:
Less than 2% for +6% - 20% variation from nominal supply voltage.
Insulation Line to ground:
2kV r.m.s.
Outputs and Set Points
Number of relays/set points:
Two.
Relay contacts:
Single pole changeover.
Rating 250V a.c. 250V d.c. max.
3A
a.c. 3A d.c. max.
Loading
(non-inductive) 750VA 30W
max.
(inductive) 750VA
3W max.
Insulation:
2kV r.m.s. contacts to earth (ground).
Number of set points:
Two.
Set point adjustment:
Programmable.
Set point hysteresis:
± 1 % fixed.
Local set point annunciation:
Red L.E.D.
Retransmission:
Number of retransmission signals:
One fully isolated.
Output current:
0 to 10mA, 0 to 20mA or 4 to 20mA programmable.
Accuracy:
±25% f.s.d. or ±0.5% reading.
Resolution:
0.1 % at 10mA, 0.05 at 20mA.
Maximum load resistance:
750 ohm (20mA max.).
Serial communication:
RS485 (optional extra). |
| |
| Mechanical
Data |
| |
Model
4620 |
|
Model
4625 |
| Mounting |
Wall |
|
Panel |
| Protection |
NEMA
4X/IP66 |
|
NEMA
4X /IP66 front |
| Dimensions |
6.36"
W x 8.43" H x 2.68" D |
|
3.78"
W x 3.78" H x 7.52" D |
| Panel
cutout |
|
|
(3.62
(+0.03/-0) x 3.62 (+0.03/-0)) |
| Weight |
4 1/2
Ibs. |
|
3 1/4
Ibs. |
|
|
|