corporateproductsnewscontact us
DeutchPolishFrench
 
 
   
back
 
       
 

dissolved oxygen analyzer spare parts Spare Parts
dissolved oxygen analyzer reagents Reagents
dissolved oxygen analyzer manuals Manuals
dissolved oxygen analyzer msds sheets MSDS Sheets

Shopping Cart

To order this complete analyzer, please call our sales office at
(800) 242-7353
 
 
Send me more information about the Dissolved Oxygen Analyzers  

µAI 4640 and µAI 4645

Dissolved Oxygen Analyzer

THIS PRODUCT IS DISCONTINUED.

The µAI-4640 wall mount and µAI-4645 panel mount Dissolved Oxygen Analyzers are user-programmable analyzers designed for dissolved oxygen measurements in a variety of applications from aeration control in sewage waste treatment and effluent monitoring to measurement of dissolved oxygen in process liquors.

 
dissolved oxygen analyzer
 
FEATURES   BENEFITS
Ultra-Reliable   No-maintenance oxygen sensor provides exceptional performance and reliability in the most demanding applications
Comprehensive Diagnostics   Sensor checking insures measurement accuracy and alerts operator of abnormal conditions
Microprocessor Flexibility   All functions are fully programmable to meet application requirements
Compact Rugged Design   Available in surface NEMA 4X or 1/4 DIN panel mount versions. When installed, transmitter front is protected to NEIVIA 4X rating
     

SENSING SYSTEMS
 
 
 
Specisifcations Floating Ball   Submersion   Rectractable
Insertion
Sensor Type D3010-166A sensor capsule   D3010-166A sensor capsule   D3010-166A sensor capsule
Response Typically 20s for 90% of a step change of oxygen concentration
at 68°F (20°C).
  Typically 20s for 90% of a step change of oxygen concentration
at 68°F (20°C).
  Typically 20s for 90% of a step change of oxygen concentration
at 68°F (20°C).
Temperature
Compensation
Automatic correction by means of an integral Pt 100 resistance thermometer.   Automatic correction by means of an integral Pt 100 resistance thermometer.   Automatic correction by means of an integral Pt 100 resistance thermometer.
Operating
Temperature
32 to 104°F (0 to 40°C)   32 to 104°F (0 to 40°C)   32 to 104°F (0 to 40°C)
Operating Pressure    Atmospheric   60 PSI   60 PSI
Minimum liquid
Velocity
30 cm/s   30 cm/s   30 cm/s
Accuracy ±0.1 ppm or ±1 % saturation within ±18°F (±10°C) of the calibration temperature in the range 32° to 95°F (0 to 35°C)
  ±0.1 ppm or ±1 % saturation within ±18°F (±10°C) of the calibration temperature in the range 32° to 95°F (0 to 35°C)   ±0.1 ppm or ±1 % saturation within ±18°F (±10°C) of the calibration temperature in the range 32° to 95°F (0 to 35°C)
Materials of
Construction
Junction Box - diecast aluminum alloy. Support arm - ABS plastic Flotation collar polypropylene   Junction Box - diecast aluminum alloy. Support arm - ABS plastic   Stainless Steel/ CPVC
Junction Box
Environmental
Protection
NEMA 4X   NEMA 4X   Retractable probe supplied with 20' of direct interconnect cable
Mounting and
Process Connections  
Pipe mounting (1 1/2" OD)   Pipe mounting (1 1/2" OD)   1 1/2" NPT male
retractable assembly
Optional mounting T adapter 1 1/2"
NPTF x 1/2" NPTF
inlet and outlet
Connection Cable Part #D1086-232. Maximum distance 3250 ft. (1000m)   Part #D1086-232. Maximum distance 3250 ft. (1000m)   Part #D1086-232. Maximum distance 3250 ft. (1000m)

Support Arm

ABS plastic OD 1 1/2". Schedule 40 for 10ft; schedule 80 for 10 to 20 ft. special   ABS plastic OD 1 1/2". Schedule 40 for 10ft; schedule 80 for 10 to 20 ft. special  

; Not applicable

 


INSTALLATION GUIDELINES
     

Display
Measured Value
:
5 digit, 7 segment back?lit LCD.

Information:
16 character, single line, dot matrix, back?lit LCD.

Ranges:
Programmable 20 to 200% saturation, 0 to 20 ppm.


Resolution:
0.1 % saturation, 0.01 ppm.

Accuracy:
±1 % saturation, 0.1 ppm.

Temperature compensation:
32° to 104°F (0° to 40°C) automatic via Pt100 resistance (RTD).

Salinity correction:
Automatic over the range 0 to 40 parts per thousand.

Environmental Data
Operating temperature limits:
4° to 131 °F (20° to 55°C).

Storage temperature limits:
13° to 158°F (25° to 70°C).

Operating humidity limits:
Up to 95% RH non?condensing.

Power Supply
Voltage requirements:
100 to 130V or 200 to 260VAC 50/60Hz.

Power Consumption:
<6VA.

Error due to power supply variation:
Less than 2% for +6% to 20% variation from nominal supply voltage.

 

Insulation:
AC to ground (line to ground) 2kV r.m.s.

Alarm Relays and Set Points
Number of relays/set points:

Two.

Set point adjustment:
Programmable.

Set point hysteresis:
±1 % of span.

Relay contacts:
Single pole changeover:
Rating     250V a.c.    250V d.c. max.
              3A a.c.        3A d.c. max.

Loading
    (non?inductive)  750VA    30W max.
    (inductive)         750VA      3W max.

Insulation:
2kV r.m.s. to ground.

Local set point annunciation:
Red L.E.D.

Retransmission
Current output:

One fully isolated, 0 to 10 mA, 0 to 20 mA or 4 to 20 mA fully programmable.

Accuracy:
±0.25% f.s.d. or ±0.5% reading.

Resolution:
0.1 % at 1 OmA, 0.05 at 20mA.
Ranges 20 to 200% saturation, 2 to 20 ppm.

Maximum load resistance:
1 k (20V max.).

Serial communication:
RS485 (optional extra)

 


 
       

 

corporate | products | service programs | news | contact us | shopping cart | home | sitemap
© 2001-2007 Waltron LLC., All rights reserved. Security & Privacy