Range:
0-20 ppb, 0-200 ppb, 0-2 ppm and 0-20 ppm with automatic range change and
four sets of normally open contacts, one of which closes for remote range
indication.
Accuracy:
=1 ppb of reading.
Precision:
±.5 ppb of reading.
Temperature Compensation:
Automatic, reduces sample temperature error to less than 0.5% of reading
per degree C.
Calibration:
Automatic on-line every 7 days or manually initiated when required.
Alarms:
Two voltage-free contacts. Alarm points set from transmitter rated at 250V
tamp, noninductive.
Dimensions:
Electronic Transmitter - 1 1.87" W x 1 1.87" H x 8" D Liquid
Handling System - 12.13" w x 16" H x 8" D Can be separated
325 feet apart. |
|
Weight:
50lbs.
Current Outputs:
Two isolated outputs - switch selectable, 0-10 mA, 0-20 mA or 4-20 mAdc,
100 ohms.
Response Time:
90% of a step change in 60 seconds.
Atmospheric Pressure
Compensation:
Programmable; preset to normal atmospheric pressure at geographical site
location.
Thermal Protection:
Automatic sample bypass if sample exceeds 131 ° F (55° C).
Power:
120/240VAC (+10%-20%) 50/60 Hz.
Degree of Protection:
Two steel cases; flow thru sample cabinet, all wetted parts stainless steel
or plastic. Finish is corrosion resistant Polyurethane. NEMA 4X. |