| PROBLEM |
|
POSSIBLE CAUSE |
|
SOLUTION |
 |
| Calibration Fail (CF) |
|
Faulty Solenoid Valve |
|
Run Solenoid Check cycle in Diagnostics – check status and connections of solenoid valve. Verify operation of valve by removing sensor from flowcell and perform another CAL with sensor exposed to atmosphere. |
 |
 |
 |
| Calibration Fail (CF) |
|
POOR (low) sensor output. |
|
Check probe handle assembly connection and other connections to sensor. |
 |
 |
 |
| Calibration Fail (CF) |
|
POOR (high) sensor output. |
|
New sensor – needs to rinse down. Let sensor run on sample for 1 hour – repeat calibration. |
 |
 |
 |
| Calibration Fail (CF) |
|
POOR (low) sensor output. |
|
Old/bad sensor. Replace sensor. |
 |
 |
 |
| Readings are not accurate – too low. |
|
Poor sensor performance. Bad calibration. Bad thermistor. |
|
Replace sensor. Check CAL log and run another CAL if last result is not good. Verify temperature as indicated on analyzer. |
 |
 |
 |
| Readings are not accurate – too high. |
|
Poor sensor performance. Bad calibration. Dissolved Oxygen leak in sample system. Bad thermistor. |
|
Let sensor run on sample water for 1-2 hours; then run another CAL. Replace sensor. Check CAL log and run another CAL if results are not good. Bench test sample water. Check temperature. |
 |
 |
 |
| Readings are not accurate – too high |
|
Defective probe handle assembly. Moisture in probe handle assembly. |
|
Measure resistance of probe handle assembly. Verify there is no voltage (mV) generated by a galvanic reaction in handle assembly. |
 |
 |
 |
| Readings are not accurate – too low |
|
Defective probe handle assembly. High resistance in assembly or corrosion on mating pins. |
|
Replace probe handle assembly. |
 |
 |
 |
| Current outputs (4-20mA) not functioning correctly. |
|
Disconnected wiring at transmitter or along wire path. DCS/recording system not set up properly. Bad CPU board. |
|
Check output set-up to make sure values are entered properly. Run Diagnostics – O/Pma Check cycle and check output coming from transmitter. Replace CPU board. |
 |
 |
 |
| Alarms not functioning correctly. |
|
Disconnected wiring at transmitter or along wire path. DCS/recording system not set up properly. Bad DIO board. |
|
Check alarm set-up to make sure values are entered properly. Run Diagnostics – Relay Check cycle and check alarm output directly from transmitter. Replace DIO board if output is not correct. |
 |
 |
 |
| Display read “HOT” |
|
Sample temperature over specified range (>131F). Faulty thermistor. |
|
Check sample temperature. Clean/replace thermistor. |
 |
 |
 |
| Display read “TEMP” |
|
No thermistor response. |
|
Clean/replace thermistor. Check thermistor connection on card. |
 |
 |
 |
| Display read “OVR” |
|
Signal from sensor too high – sample concentration over maximum range (>20ppm) |
|
Check sample concentration. Check sensor connections. |
 |
 |
 |