Range:
User configurable to any two consecutive decades of concentration
between 0.05 and 5000 ppm depending on analysis.
Repeatability:
±2% of reading.
Reproducibility:
±3% of reading.
Response Time:
Less than 5 minutes for 90% step change.
Millivolt Range:
-1000 to + 1000mV.
Millivolt Resolution:
±0.1 mV.
Control Temperature Range:
30 to 45°C.
Temperature Resolution:
±0.1 °C.
Displays:
Concentration: -5 digit blue fluorescent
Information: -20 character dot matrix blue fluorescent
Status Indication:
Two flashing LEDs in alarm state.
Single LED lit when 'HOLD' switch operated.
Single LED lit when calibration is in progress.
Single LED lit when the monitor is 'Out of Service'.
Current Outputs:
As standard one isolated current output 0-1, 0-10, 0-20, or 4-20mA
selectable via plug-in links.
Current Output Span:
Any 1 to 2 decades of display range, ogarithmic or linear.
|
|
Computer Interface:
Optional second current output or RS422/RS423 serial interface.
Alarms:
Two high or low concentration alarms. Remote calibration in progress
indication. Remote 'Out of Service' indication. All voltage free
250V, 5A non-inductive relay contacts.
Concentration Alarm Adjustment:
Programmable over assigned range.
Concentration Alarm Differential:
0 to 5%.
Concentration Alarm Delay:
0 to 60 minutes.
Programmable Date Retention:
3 years.
Calibration:
Fully automatic two point, plus manual initiation on demand.
Routine Maintenance:
Once a month: Replenish reagents, clean flow system. Once a year:
Replace pump tubing and pump capstans.
Power Supply:
110-120V or 220-240V, 50/60Hz, 100VA.
Power Supply Tolerance:
+6% to -20%.
Isolation Voltage:
Input, output and power supply 1.5kV.
Dimensions:
Height: 34-3/4" (883mm)
Width: 20-9/10" (531 mm)
Depth: 7-4/5" (199mm)
Degree of Protection:
Electronics Section: NEMA 4X.
Liquid Handling: Critical internal components NEMA 4X. |