AQUALYZER®
9001 SERIES
Hydrazine Analyzer
Aqualyzer® 9071
The Waltron Aqualyzer® 9071 Hydrazine Analyzer is the most advanced hydrazine instrument specifically developed for use in continuous measurement and determination of hydrazine concentration in boiler feedwater.
Common Hydrazine Analyzer application points in a boiler system for online hydrazine measurement include:deaerator, condensate system and boiler feedwater.
The Aqualyzer® 9071 hydrazine analyzer operates automatically and comes complete with true fully automatic calibration.
Download Hydrazine Analyzer 9071 brochure (PDF)
|
|
|
| FEATURES |
|
BENEFITS |
 |
 |
 |
 |
| Wide analysis range: 0.01ppb – 1ppm |
|
Low reagent and standard consumption |
 |
 |
 |
| True fully automatic 1-point calibration |
|
Minimal maintenance required |
 |
 |
 |
| Low maintenance: autonomous instrument for up to two months |
|
Automatic operation |
 |
 |
 |
| Grab sample capability |
|
|
 |
 |
 |
| No change of diffusion tubing |
|
|
 |
 |
 |
| Automatic range selection |
|
|
 |
 |
 |
| Process calibration – calibrate to sample |
|
|
 |
 |
 |
| Two 4-20 mA isolated current outputs |
|
|
 |
 |
 |
| Protection of sensor from “Hot Sample” |
|
|
 |
 |
 |
| Automatic temperature compensation |
|
|
 |
 |
 |
| Automatically stores last 10 calibrations and alarm logs |
|
|
 |
 |
 |
| Rechargeable hydrazine cell |
|
|
 |
 |
 |
Aqualyzer® 9071 Sensor Unit
During operation,the sample flows through the heat exchanger,passes through constant head (used to stabilize minor changes in flow rate)and through the two solenoid valves. A vapor entrainment mechanism adds alkaline buffer vapor to the sample before it enters the hydrazine cell and flows to drain. Calibration and grab sample are controlled automatically by the analyzer. Standard/grab sample liquid is brought into system via the two 3-way solenoid valves. A thermistor is housed in the sensor and is used to detect the temperature of the sample/standard/grab sample.
|
|
|
| |
| |
| Operating
Conditions |
 |
 |
 |
| Sample |
|
Temperature: 41 – 131° F (5 – 55° C); Flow 150 – 400 ml/min |
 |
 |
 |
| Ambient
Temperature |
|
32 – 131° F (0 – 55° C) |
 |
 |
 |
| Pressure |
|
5–30 psig (0.35 – 2.07 bar) |
 |
 |
 |
| Composition |
|
Sample should be filtered to 60 microns, free of film forming compounds |
 |
 |
 |
|
|