µAI 9043

Silica / Phosphate Analyzer

REVOLUTIONARY PATENTED TECHNOLOGY

Our Colorimetric Silica/Phosphate Analyzer (Aqualyzer® Series 9040) is based on the new patented Loop Flow Analysis (LFA) cycle.

Common application points in a boiler system for online silica measurement include:make-up water,condensate,boiler feedwater,saturated steam/main steam and cooling water. Application points for phosphate include: boiler water and cooling water.

LFA technology involves isolating,manipulating and testing a known volume of sample in a closed loop system. A powerful programmable microprocessor allows for customized program cycles,ensuring accuracy and reliability with every application.

Download Silica / Phosphate Analyzer 9043 brochure (PDF)

 


µAI 9040 Series - Cilica / Phosphate Analyzer

Please contact us
for panel mount version.

 
FEATURES   BENEFITS
Highest accuracy (e.g.for silica:+/– 1ppb or +/–2% of reading)   Low reagent and standard consumption
Widest range analysis (e.g.for silica: ranges available from 0.1ppb to 200ppm)   Minimal maintenance required
Single, dual stream or multi-parameters SiO2/PO4   Automatic operation
Short cycle times with no sample heater needed    
Automatic dilution when above normal concentration range    
Automatic calibration    
Adjustable cycle and calibration frequency    
Ultra-low reagent consumption and smallest size instruments    
No reagent pumps or compressed air requirements    
Grab sample capability    
Standardized flow through constant head with automatic flow detection    
Last 400 results stored internally    
     

GENERAL SPECIFICATIONS
Performance
Sensor Classification   Colormetric with glass flowcell and LED source
Application   Demineralized,boiler,potable,surface and waste water
Power Requirements   12V
Power Supply   110-120V or 220-240V 50/60 Hz, 100 VA
Humidity   Up to 90% not condensable
Ambient Temperature Range   10-45° C analyzer (50-113° F)
Range   Silica: 0-500 ppb, 0-5000 ppb, 0-200 ppm*
Phosphate: 0-2.4 ppm,0-7 ppm, 0-20 ppm, 0-200 ppm*
Hydrazine, Copper: 0-100 ppb, 0-1 ppm*
Ammonia: 0-750 ppb, 0-7.5 ppm*
Ethylene Glycol: 0-15 ppm, 0-150 ppm*
*By Auto Dilutio
Accuracy   Silica: +/– 1 ppb or +/– 2% of reading, whichever is greater
Phosphate, Hydrazine, Ammonia, Copper, Ethylene Glycol: +/- 2% of full scale
Unit Dimensions   Height = 31.5" (80.01 cm), Width = 16.54" (42.01 cm), Depth = 11.13" (28.27 cm)
Atmospheric Pressure Range   No Limits
Effect of Electromagnetic Fields   EMC tested according to CE compliance
Tolerance to Electrostatic Discharges   EMC tested according to CE compliance
Positioning and
installation details
  Wall or panel mounting analyzer,installed approx 39" (99cm) from the ground,maximum 13’ (396cm) from sampling point
Response Time (approximate)   Silica: 10 min, Phosphate: 6 min, Hydrazine: 12 min, Ammonia: 12 min, Copper: 9 min, Ethylene Glycol: 10 min
Alarms   Concentration, cal error, out of service/sample
Reproducibility   Silica:+/– 1 ppb or +/– 2% of reading, whichever is greater
Phosphate, Hydrazine, Ammonia, Copper, Ethylene Glycol: +/– 2% of reading
Degree of Protection   IP55 (NEMA 4): Wet section, IP65 (NEMA 4x): Electronics
Required Maintenance   Once a month replenish reagents, once a year replace tubing
Reagent Consumption (approximate)   Silica, Phosphate, Ammonia, Copper: 1 liter each per month
Hydrazine, Ethylene Glycol: 4 liters each per month
Reactor Volume   10ml
Materials in Contact with Sample   Glass, Silicone, Plexiglass, Stainless Steel AISI 316
CE Compliance   Meets low voltage and low electromagnetic compatibility directives
Sample Conditioning Requirements   Filtering between 10 and 60 micron, depending on the matrix, needed only to avoid clogging. Internal disposable filters supplied (60 micron)
     
Sample Delivery Operating Conditions
Temperature Range   5-55°C (41-131° F)
Flow   150-200 ml/min, 5-30 psig (0.35 – 2.07 bar)
Turbidity   Not applicable,sample blank correction
Color   Not applicable; sample blank correction
PH   3-12
     
Signal Outputs
4-20 mA, Voltage 0-5   4-20mA (galvanic isolator module available as option) or 0-5V
Printer Options   Optional, serial output RS232 or 485
Serial I/O for Signals   Serial data output RS232 or 485 available as option
     
Operational Calibration
Frequency/Intervals   Recommended:7days
Single/Multi Point   Multi-Point:zero and mid range
Matrix Corrections   Yes,sample blank correction
Manual/Automatic   Both
     
 
       
 