| FEATURES |
|
BENEFITS |
 |
| Highest accuracy (e.g.for silica:+/– 1ppb or +/–2% of reading) |
|
Low reagent and standard consumption |
 |
 |
 |
| Widest range analysis (e.g.for silica: ranges available from 0.1ppb to 200ppm) |
|
Minimal maintenance required |
 |
 |
 |
| Single, dual stream or multi-parameters SiO2/PO4 |
|
Automatic operation |
 |
 |
 |
| Short cycle times with no sample heater needed |
|
|
 |
 |
 |
| Automatic dilution when above normal concentration range |
|
|
 |
 |
 |
| Automatic calibration |
|
|
 |
 |
 |
| Adjustable cycle and calibration frequency |
|
|
 |
 |
 |
| Ultra-low reagent consumption and smallest size instruments |
|
|
 |
 |
 |
| No reagent pumps or compressed air requirements |
|
|
 |
 |
 |
| Grab sample capability |
|
|
 |
 |
 |
| Standardized flow through constant head with automatic flow detection |
|
|
 |
 |
 |
| Last 400 results stored internally |
|
|
 |
 |
 |
| |
|
|
| GENERAL SPECIFICATIONS |
 |
| Performance |
 |
 |
 |
| Sensor
Classification |
|
Colormetric with glass flowcell and LED source |
 |
 |
 |
| Application |
|
Demineralized,boiler,potable,surface and waste water |
 |
 |
 |
| Power Requirements |
|
12V |
 |
 |
 |
| Power Supply |
|
110-120V or 220-240V
50/60 Hz, 100 VA |
 |
 |
 |
| Humidity |
|
Up to 90% not
condensable |
 |
 |
 |
| Ambient Temperature
Range |
|
10-45° C analyzer
(50-113° F) |
 |
 |
 |
| Range |
|
Silica: 0-500 ppb, 0-5000 ppb, 0-200 ppm*
Phosphate: 0-2.4 ppm,0-7 ppm, 0-20 ppm, 0-200 ppm*
Hydrazine, Copper: 0-100 ppb, 0-1 ppm*
Ammonia: 0-750 ppb, 0-7.5 ppm*
Ethylene Glycol: 0-15 ppm, 0-150 ppm*
*By Auto Dilutio |
 |
 |
 |
| Accuracy |
|
Silica: +/– 1 ppb or +/– 2% of reading, whichever is greater
Phosphate, Hydrazine, Ammonia, Copper, Ethylene Glycol: +/- 2% of full scale |
 |
 |
 |
| Unit
Dimensions |
|
Height = 31.5" (80.01 cm), Width = 16.54" (42.01 cm), Depth = 11.13" (28.27 cm) |
 |
 |
 |
| Atmospheric
Pressure Range |
|
No Limits |
 |
 |
 |
| Effect of
Electromagnetic Fields |
|
EMC
tested according to CE compliance |
 |
 |
 |
| Tolerance
to Electrostatic Discharges |
|
EMC
tested according to CE compliance |
 |
 |
 |
Positioning
and
installation details |
|
Wall or panel mounting analyzer,installed approx 39" (99cm) from the ground,maximum 13’ (396cm) from sampling point |
 |
 |
 |
| Response Time (approximate) |
|
Silica: 10 min, Phosphate: 6 min, Hydrazine: 12 min, Ammonia: 12 min, Copper: 9 min, Ethylene Glycol: 10 min |
 |
 |
 |
| Alarms |
|
Concentration,
cal error, out of service/sample |
 |
 |
 |
| Reproducibility |
|
Silica:+/– 1 ppb or +/– 2% of reading, whichever is greater
Phosphate, Hydrazine, Ammonia, Copper, Ethylene Glycol: +/– 2% of reading |
 |
 |
 |
| Degree of
Protection |
|
IP55 (NEMA 4): Wet section, IP65 (NEMA 4x): Electronics |
 |
 |
 |
| Required Maintenance |
|
Once a month
replenish reagents, once a year replace tubing |
 |
 |
 |
| Reagent Consumption (approximate) |
|
Silica, Phosphate, Ammonia, Copper: 1 liter each per month
Hydrazine, Ethylene Glycol: 4 liters each per month |
 |
 |
 |
| Reactor Volume |
|
10ml |
 |
 |
 |
| Materials
in Contact with Sample |
|
Glass,
Silicone, Plexiglass, Stainless Steel AISI 316 |
 |
 |
 |
| CE Compliance |
|
Meets low voltage and low electromagnetic compatibility directives |
 |
 |
 |
| Sample Conditioning
Requirements |
|
Filtering between 10 and 60 micron, depending on the matrix, needed
only
to avoid clogging. Internal disposable filters supplied (60 micron) |
 |
 |
 |
| |
|
|
| Sample
Delivery Operating Conditions |
 |
 |
 |
| Temperature
Range |
|
5-55°C (41-131° F) |
 |
 |
 |
| Flow |
|
150-200 ml/min, 5-30 psig (0.35 – 2.07 bar) |
 |
 |
 |
| Turbidity |
|
Not applicable,sample blank correction |
 |
 |
 |
| Color |
|
Not applicable;
sample blank correction |
 |
 |
 |
| PH |
|
3-12 |
 |
 |
 |
| |
|
|
| Signal
Outputs |
 |
 |
 |
| 4-20 mA, Voltage
0-5 |
|
4-20mA (galvanic
isolator module available as option) or 0-5V |
 |
 |
 |
| Printer Options |
|
Optional, serial
output RS232 or 485 |
 |
 |
 |
| Serial I/O
for Signals |
|
Serial data output RS232 or 485 available as option |
 |
 |
 |
| |
|
|
| Operational
Calibration |
 |
 |
 |
| Frequency/Intervals |
|
Recommended:7days |
 |
 |
 |
| Single/Multi
Point |
|
Multi-Point:zero and mid range |
 |
 |
 |
| Matrix Corrections |
|
Yes,sample blank correction |
 |
 |
 |
| Manual/Automatic |
|
Both |
 |
 |
 |
| |
|
|