µAI 9030 SERIES

Sodium Analyzer

THIS PRODUCT IS DISCONTINUED and has been replaced with our NEW Sodium Analyzer Aqualyzer® 9031 and the compact version 9031C.

The Waltron µAI 9030 Sodium Analyzer is a microprocessor-based instrument specifically developed for use in continuous measurement and determination of sodium ion concentration. The µAI 9030 Sodium Analyzer is designed to meet the exacting standards for low level sodium measurement of nuclear and fossilfueled power stations, providing fast response, high measurement accuracy and microprocessor programming flexibility of all control output functions.

 

 
FEATURES    
• Wide Analysis Range (0.01 PPB - 10 PPM)   • Self Diagnostics
• Automatic Range Selection   • Low Maintenance
• Vapor Entrainment Buffering   • Dual Isolated Current Outputs
• Automatic Two Point Calibration   • Dual Programmable Alarms

µAI 9030 Sensor Unit
During calibration the 3-way valve is activated, a standard of known sodium concentration passes through a heat exchanger to equilibriate the solution's temperature to that of the sample. The standard then enters the constant head unit where flow is stabilized. Alkaline buffer vapor is added via a vapor entrainment mechanism before the standard enters the sodium measuring flow cell and flows to drain. During normal operating conditions the standard solution is replaced by the sample.



GENERAL SPECIFICATIONS
Performance
Range   Linear: 0-1 ppb, 0-10 ppb, 0-.1 ppm, 0-1 ppm, 0-10 ppm.
Logarithmic: .01-1 ppb, .1-10 ppb, .001-.1 ppm,.01-1 ppm, .1-10 ppm.
Selectable: Linear/ Logarithmic, with four sets normally open contacts for remote range indication.
Accuracy   +/- 5% of reading or +/- .1 ppb (whichever is greater) within +/- 5°C of calibration temperature.
Reproducibility   +/- 5% of reading or +/- .1 ppb (whichever is greater) at constant temperature.
Response Time   90% of 1-10 ppb step: less than 4 min.
90% of 100-1 ppb step: less than 6 min.
Temperature Compensation   Automatic.

pH Compensation

  Alkaline vapor buffer:
0.88 SG (35% W/V) ammonia above 0.1 ppb sodium.
50% diethylamine solution below 0.1 ppb sodium.
Calibration   Automatic sequence, manually initiated by operator every one to four weeks.
Outputs   Two isolated outputs - switch selectable, 0-10 mA, 0-20 mA or 4-20 mAdc, 1000 ohms maximum.
Alarms   High & low voltage free contacts rated 2A, 250 VAC non-inductive.
Power   120/240VAC (+10%-20%) 50/60 Hz.
 
 
Operating Conditions
Sample   Temperature: 41 - 131 °F (5 - 55°C); Flow 50 - 500 ml/min.
Ambient Temperature   32 - 131 °F (0 - 55°C).
Pressure   5 - 100 psig.
Composition   Filter to 60 micron, free of film forming compounds.
 
 
Mechanical
Construction   Two steel cases; flow thru sample cabinet; all wetted parts stainless steel or plastic.
Dimensions   Indicating Transmitter - Width: 11.87", Height: 11.87", Depth: 8.00".
Sample Cabinet - Width: 12.13", Height: 16.00", Depth: 8.00".
Mounting   Indicating Transmitter - Four holes: 5/16" diameter, Horizontal: 9.06", Vertical: 9.06"
Sample Cabinet - Four holes: 5/16", Horizontal: 9.06", Vertical: 13", 10" clearance desirable above unit for calibration bottles.
Weight   Indicating Transmitter - 25 lbs; Sample Cabinet - 25 lbs.
Wiring   Six (6) 1 /2" conduit knockouts.
Piping   Sample Inlet: 1 /4" OD swagelok.
Drain: 3/8" OD tube.
Finish   Corrosion resistant Polyurethane. Maximum Separation.
Distance   325 feet (100 meters).
 
       

 