Sodium Analyzer
THIS PRODUCT IS DISCONTINUED and has been replaced with our NEW Sodium Analyzer Aqualyzer® 9031 and the compact version 9031C.
The Waltron µAI
9030 Sodium Analyzer is a microprocessor-based instrument specifically
developed for use in continuous measurement and determination of sodium
ion concentration. The µAI 9030 Sodium Analyzer is designed to meet the exacting standards
for low level sodium measurement of nuclear and fossilfueled power stations,
providing fast response, high measurement accuracy and microprocessor programming
flexibility of all control output functions. |
|
|
| FEATURES |
|
|
 |
| Wide Analysis
Range (0.01 PPB - 10 PPM) |
|
Self Diagnostics |
 |
 |
 |
| Automatic Range
Selection |
|
Low Maintenance |
 |
 |
 |
| Vapor Entrainment
Buffering |
|
Dual Isolated
Current Outputs |
 |
 |
 |
| Automatic Two
Point Calibration |
|
Dual Programmable
Alarms |
 |
 |
 |
µAI 9030 Sensor Unit
During calibration the 3-way valve is activated, a standard of known sodium
concentration passes through a heat exchanger to equilibriate the solution's
temperature to that of the sample. The standard then enters the constant
head unit where flow is stabilized. Alkaline buffer vapor is added via a
vapor entrainment mechanism before the standard enters the sodium measuring
flow cell and flows to drain. During normal operating conditions the standard
solution is replaced by the sample.

| GENERAL
SPECIFICATIONS |
 |
| Performance |
 |
 |
 |
| Range |
|
Linear:
0-1 ppb, 0-10 ppb, 0-.1 ppm, 0-1 ppm, 0-10 ppm.
Logarithmic: .01-1 ppb, .1-10 ppb, .001-.1 ppm,.01-1 ppm, .1-10 ppm.
Selectable: Linear/ Logarithmic, with four sets normally open contacts for
remote range indication. |
 |
 |
 |
| Accuracy |
|
+/- 5%
of reading or +/- .1 ppb (whichever is greater) within +/- 5°C of calibration
temperature. |
 |
 |
 |
| Reproducibility |
|
+/- 5%
of reading or +/- .1 ppb (whichever is greater) at constant temperature. |
 |
 |
 |
| Response
Time |
|
90% of
1-10 ppb step: less than 4 min.
90% of 100-1 ppb step: less than 6 min. |
 |
 |
 |
| Temperature
Compensation |
|
Automatic. |
 |
 |
 |
|
pH Compensation
|
|
Alkaline
vapor buffer:
0.88 SG (35% W/V) ammonia above 0.1 ppb sodium.
50% diethylamine solution below 0.1 ppb sodium. |
 |
 |
 |
| Calibration |
|
Automatic
sequence, manually initiated by operator every one to four weeks. |
 |
 |
 |
| Outputs |
|
Two isolated
outputs - switch selectable, 0-10 mA, 0-20 mA or 4-20 mAdc, 1000 ohms maximum. |
 |
 |
 |
| Alarms |
|
High & low voltage
free contacts rated 2A, 250 VAC non-inductive. |
 |
 |
 |
| Power |
|
120/240VAC
(+10%-20%) 50/60 Hz. |
 |
 |
| |
| |
| Operating
Conditions |
 |
 |
 |
| Sample |
|
Temperature:
41 - 131 °F (5 - 55°C); Flow 50 - 500 ml/min. |
 |
 |
 |
| Ambient
Temperature |
|
32 -
131 °F (0 - 55°C). |
 |
 |
 |
| Pressure |
|
5 - 100
psig. |
 |
 |
 |
| Composition |
|
Filter
to 60 micron, free of film forming compounds. |
 |
 |
 |
| |
| |
| Mechanical |
 |
 |
 |
| Construction |
|
Two steel
cases; flow thru sample cabinet; all wetted parts stainless steel or plastic. |
 |
 |
 |
| Dimensions |
|
Indicating
Transmitter - Width: 11.87", Height: 11.87", Depth: 8.00".
Sample Cabinet - Width: 12.13", Height: 16.00", Depth: 8.00". |
 |
 |
 |
| Mounting |
|
Indicating
Transmitter - Four holes: 5/16" diameter, Horizontal: 9.06", Vertical:
9.06"
Sample Cabinet - Four holes: 5/16", Horizontal: 9.06", Vertical:
13", 10" clearance desirable above unit for calibration bottles. |
 |
 |
 |
| Weight |
|
Indicating
Transmitter - 25 lbs; Sample Cabinet - 25 lbs. |
 |
 |
 |
| Wiring |
|
Six (6)
1 /2" conduit knockouts. |
 |
 |
 |
| Piping |
|
Sample
Inlet: 1 /4" OD swagelok.
Drain: 3/8" OD tube. |
 |
 |
 |
| Finish |
|
Corrosion
resistant Polyurethane. Maximum Separation. |
 |
 |
 |
| Distance |
|
325 feet
(100 meters). |
 |
 |
 |
|
|