AQUALYZER®
9001 SERIES
Sodium Analyzer
Aqualyzer® 9031
The Waltron Aqualyzer® 9031 Sodium Analyzer is the most advanced online instrument specifically developed for use in continuous measurement and determination of sodium ion concentration.
Common Sodium Analyzer 9031 application points in a boiler system for online sodium measurement include:make-up water,condensate,boiler feedwater and saturated steam/main steam.
The Aqualyzer® 9031 sodium analyzer operates automatically and comes complete with true fully automatic calibration.
Download Sodium Analyzer 9031 brochure (PDF)
|
|
|
| FEATURES |
|
BENEFITS |
 |
 |
 |
 |
| Wide analysis range: 0.01ppb – 1ppm |
|
Low reagent and standard consumption |
 |
 |
 |
| True fully automatic single or 2-point calibration |
|
Minimal maintenance required |
 |
 |
 |
| Low maintenance: no change of standards for up to two months (one month for reagent) |
|
Automatic operation |
 |
 |
 |
| Grab sample capability |
|
|
 |
 |
 |
| No change of diffusion tubing |
|
|
 |
 |
 |
| Automatic range selection |
|
|
 |
 |
 |
| Process calibration – calibrate to sample |
|
|
 |
 |
 |
| Two 4-20 mA isolated current outputs |
|
|
 |
 |
 |
| Protection of electrodes from “Hot Sample” |
|
|
 |
 |
 |
| Automatic temperature compensation |
|
|
 |
 |
 |
| Automatically stores last 10 calibrations and alarm logs |
|
|
 |
 |
 |
| No use of highly toxic reagents such as monoethylamine or diisopropylamine |
|
|
 |
Aqualyzer® 9031 Sensor Unit
The sample flows through the heat exchanger, passes through
constant head (used to stabilize minor changes in flow rate)
and through the 3 solenoid valves. A vapor entrainment
mechanism adds alkaline buffer vapor to the sample
before it enters the sodium measuring flow cell and flows to drain.
Calibration and grab sample are controlled automatically
by the analyzer. Standard/grab sample liquid is brought into
system via the three 3-way solenoid valves. A thermistor is
housed in the flowcell and is used to detect the temperature
of the sample/standard/grab sample.
|
|
 |
 |
 |
| |
| |
| Operating
Conditions |
 |
 |
 |
| Sample |
|
Temperature:41 – 131°F (5 - 55°C); Flow 150 – 400 ml/min |
 |
 |
 |
| Ambient
Temperature |
|
32 – 131°F (0 – 55°C) |
 |
 |
 |
| Pressure |
|
5 – 30 psig (0.35 – 2.07 bar) |
 |
 |
 |
| Composition |
|
Sample should be filtered to 60 microns, free of film forming compounds |
 |
 |
 |
|
|