µAI 4670 and µAI 4675

Turbidity Analyzer

THIS PRODUCT IS DISCONTINUED.

A Waltron Turbidity system is comprised of a 4670 wallmounted or 4675 panel-mounted analyzer, together with one of four sensors Model 7997-201 to 7997-401 .

The sensors are available in flow and dip versions. Where appropriate they are supplied with auto-cleaning to minimize maintenance on more turbid samples.

 

 
Turbidity Analyzers 4670 and 4675

FEATURES   BENEFITS
Sensing System Flexibility   Operator can choose four sensor cell systems utilizing flowthru or dip configurations for a wide range of applications.
Comprehensive Self-Diagnostics   Ensures measurement accuracy by automatically checking sensor performance and ensuring that the light source is operating within the specification.
Ultra-Low Back Scatter   Enables true zero setting to ensure accurate and reliable results below 0.1 NTU.
Dry Secondary Calibration Standard   Simplifies accurate and repeatable calibration and eliminates the need and cost of chemical standards.
Auto-Cleaning Systems   Maintains superb performance and reduces maintenance by a programmable wiper cleaning system.
Lifetime Zero   Ensures high accuracy, minimum maintenance and long-term performance.
Wide Analysis Range   Programmed to read out in either NTU or FTU from 0-5 NTU to 0-1000 FTU. Ideal for applications varying from final treatment of potable water to effluent discharge monitoring.
Advanced Photodiode Technology   Extremely low noise signal enabling high accuracy and resolution to 0.01 NTU. Essential for accurate results below 1.0 NTU.
     
SENSOR SELECTION GUIDE
       
Specisifcations      7997-211
Turbidity Sensor
  7997-201
Turbidity Sensor
  7997-301
Turbidity Sensor
Range: Programmable 0 - 1NTU to 0 - 30 NTU   Programmable 0 - 50 NTU to 0 - 250 NTU   Programmable 0 - 100 FTU to 0 - 500 FTU and 0 to 1000 mg/I (or ppm)
Principle: 90° scattered light measurement. Compliant to IS07027   90° scattered light measurement. Compliant to IS07027   Transmitted light
Characteristic: Linear based on Formazine   Linear based on Formazine   Logarithmic based on Formazine
Resolution: 0.01 NTU   0.1 NTU   1 FTU Repeatability
Better than 1 % of span
Repeatability: Better than 1 % of span   Better than I % of span   Better than I % of span
Accuracy: ±2% FSD   ±2% FSD   ±2% FSD
Temperature Drift: 0.005 NTU/10° C   0.005 NTU/10° C   0.2 FTU/° C
Response Time:
varies with flow rate, typically 90% step change in less than 45s at 1 l/min   Varies with flow rate, typically 90% step change in 2 minutes at 1 l/min   Varies with flow rate, typically 90% step change in 2 minutes at 1 l/min
Flow Rate: 0.5 I to 1. 5 l/min   0.5 I to 1.5 l/min   0.5 I to 1.5 l/mm
Sample Operating
Temperature:
  
0 to 50° C   0 to 50°C   0 to 50°C
Sample Pressure: Up to 3 bar   Up to 3 bar   Up to 3 bar

Integral Wiper
Cleaning System:

    AutoCleaning
Programmable operational frequency every 0.25 hour, 0.5 hour, 0.75 hour or mulitples of 1 hour up to 24 hours
  AutoCleaning
Programmable operational frequency every 0.25 hour, 0.5 hour, 0.75 hour or mulitples of 1 hour up to 24 hours

Applications:

Best suited for final treated potable and filtered water.   Best suited for raw water on potable water treatment plants. Includes auto-deaning.   Best suited for effluent discharge monitoring, water intakes and flier backwash. Includes auto-cleaning.
  7997-401
Best suited for open channels and tanks and effluent discharge monitoring. Includes auto-cleaning.

  


GENERAL SPECIFICATIONS
     
DISPLAY
Measured value:

5-digit x 7-segment backlit LCD

Information:
16-character, single line, dot matrix, backlit LCD

Ranges:
Used with 7997-10 1 0 - Z to 0 - 30 NTU
Used with 7997-201 0 - 50 to 0 - 250 NTU
Used with 7997-301 0 - 100 to 0 - 500 FLU and 0 - 1000 mg/I
Used with 7997-401 0 - 100 to O - 1000 FTU and 0 - 2000 mg/l

Units of measurement:
NTU and FTU all models mg/l on models 7997-301 and 7997-401

Accuracy:
=0.2% F.s.d., ±1 digit

Linearity:
t0.1 % f.s.d.

Auto-clean timing:
Programmable 15 min, 30 min, 45 min or 1 hour up to 24 hours in I hour increment

ENVIRONMENTAL DATA
Operating temperature limits:

200 to 550 C (-4° to 131 0 F)

Storage temperature limits:

250 to 55° C (-130 to 131 ° F)

Operating humidity limits:
Up to 95% RH non-condensing

POWER SUPPLY
Voltage requirements:

100 to 130V, 200 to 260V, 50/60Fiz

Power consumption:
<bVA a.c.

Error due to power supply variation:
Less than 0.1 % for +6% to -20% variation from nominal supply

Insulation:
Line to ground 2kV r.m.s.
   

  RELAY OUTPUTS AND SET POINTS
No. of relays:

Two

Relay contacts:
Single pole changeover
Rating 250V a.c., 250V d.c. max.
           3Aa.c. 3Ad.c. max.
Loading (non-inductive) 750VA 30W max.
             (inductive) 750VA 3W max.

Insulation:
2kV r.m.s. contacts ground

No. of set points:
Two

Set point adjustment:
Programmable

Set point hysteresis:
-1 % fixed

Local set point annunciation:
Red L. E. D.

RETRANSMISSION
No. of retransmission signals:

One fully isolated programmable 0 - 1 OmA, 0 - 20mA or 4 - 20mA

Accuracy:
10.25% f.s.d. -0.5% reading

Resolution:
0.1 % at I OmA, 0.05% at 20mA
Max. load resistance 75052 (20mA max.)


MECHANICAL DATA
  Model 4670    Model 4675
Mounting    wall mounting      panel mounting   
Protection    IP65 IP66 front
Weight    4.5Ibs 3.25Ibs
Dimensions    (6.30 in) wide    (3.78 in) wide   
  (8.43 in) high (3.78 in) high
  (2.68 in/ deep (7.52 in) deep

 
   
 
 